Customization: | Available |
---|---|
After-sales Service: | Yes |
Power Supply: | Yes |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Product name:
Rapid Temperature Change test Chamber
Product Description
This test chamber is suitable for high and low temperature tests and rapid temperature change tests of finished products and components, so as to evaluate the performance and quality of the test samples. It has the following several basic characteristics:
The structure design of the test system is advanced and reasonable, the manufacturing process is standardized, the appearance is beautiful and elegant, and the maintenance is convenient.
2. The matching and assembly of components are well-matched. The main functional components all adopt original imported parts with international advanced levels, which enhances the safety and reliability of the product and can meet the users' requirements for long-term and high-frequency use.
3. The necessary conditions for the normal operation of the equipment have been provided. The buyer only needs to offer power and electricity, test consumables (such as distilled water or purified water), and a working environment that meets the requirements.
Conform to the standard
1.GB10592.2-1989 Technical Conditions for High and Low Temperature Test Chambers
2.GB11158-1989 Technical Conditions for High Temperature Test Chambers
3.GB105589-1989 Technical Conditions for Low Temperature Test Chambers
4.GB10586-1989 Technical Conditions for Wet Heat Test Chambers
5.GB/T2423.1-2001 (IEC68-2-1) Test A: Low Temperature Test Method
6.GB/T2423.2-2001 (IEC68-2-2) Test B High Temperature Test Method
7.GJB150.3-1986 High-temperature test
8.GJB150.4-1986 Low-temperature Test
9.GB/T2423.3-2001 Test Method for Constant Damp Heat
10.GB2423.22 Test Method for Temperature Variation
11.GBT2423.4-1993 Test Method for Alternating Wet Heat
Technical parameters
Humidity fluctuation degree | ≤±3%R.H |
Humidity deviation | ≤±3%R.H (Note: ≤±5%R.H below 75%R.H) |
Internal dimensions | 500*600*500mm (W*H*D) |
External dimensions | Subject to the actual size |
Temperature range | -20ºC(-40ºC, -60ºC, 0ºC, 20ºC) to 150ºC |
Humidity range | 20% to 98%R.H (Relative humidity) |
Heating and cooling time | Linear range: -30ºC to +125ºC, 7 to 10ºC per minute; +125ºC to -30ºC, linear range: 7 to 10ºC per minute |
Temperature fluctuation degree | Plus or minus 0.5 ºC or less |
Temperature deviation | Plus or minus 2 ºC or less |
Product features:
Main configuration:
Box:
Structural form: Integrated type, with the test chamber, controller and refrigeration unit installed separately
Shell material: Cold-rolled steel plate with baked paint
Studio material: Stainless steel plate SUS304
Insulation material: Rigid polyurethane foam, insulation layer thickness 100mm
Condensate outlet hole: It has an outlet hole for the condensate water of the unit
Test hole: A Φ50 test hole is opened at the center of the left side of the box
Observation window: A 240*320mm hollow resistance film heating anti-frost observation window
Lighting device: 1 energy-saving lamp
Control System:
Temperature controller: TEMI880 programmable liquid crystal touch controller
Temperature control performance: Temperature resolution: ±0.1ºC;
Temperature sensor: A-level high-precision industrial platinum thermal resistance PT100;
Operation mode: Program and constant (fixed point);
Main actuating appliances: "LG" AC contactor, thermal relay; "Omron" small relay Transformer "Xurui" solid state relays, etc.
Installation Features:
A. The entire refrigeration system was filled with nitrogen and maintained at a pressure of 2.2Mpa for leak detection test.
B. The heating and cooling systems are completely independent;
C. Equipped with the company's feature, the high-temperature refrigeration machine return gas cooling system;
D. All the action programs of the refrigeration system are fully automatically controlled by the instruments.
E. Heating system: Temperature control elements, solid-state relays;
F. Heating method: High-quality nickel-chromium alloy wire electric heater;
G Blower system, circulation system: Fan forced convection method;
H. Motor: Extended stainless steel shaft motor;
I. Fan blade: Centrifugal fan blade;
J. Vibration reduction measures: Compressor, special rubber vibration reduction.
Safety device:
1. Safety alarm action: Sound and light alarm prompts;
2. Test chamber structural components: Grounding protection;
3. Test chamber chamber: Independent over-temperature protection system;
4. Compressor: Overpressure protection, over-temperature protection, overload protection;
5. Power supply: Overcurrent/leakage protection;
6. Random accessories: 1 power cord;